W-C-596H/GEN
TABLE IV. Qualification retention for hospital grade connectors - Continued.
Inspection |
Plug |
Receptacle |
Cable outlet |
Male inlet |
Requirement paragraph |
Test paragraph |
Group VII (6 samples) 1/ Subgroup A (3 samples) 1/ Conditioning (UL 498) (.204 pin) Grounding pin retention (UL 498) Plug connection and separation test (UL 498) Visual and mechanical Subgroup B (3 samples) 1/ Conditioning (UL 498) (with plug) Grounding contact temperature (UL 498) Grounding path resistance (UL 498) Visual and mechanical |
x x x x x x x x |
3.6.14g 3.6.14e 3.6.14m 3.1, 3.4, 3.5, 3.7 and 3.8 3.6.14g 3.6.14j 3.6.14n 3.1, 3.4, 3.5, 3.7 and 3.8 |
4.6.15g 4.6.15e 4.6.15m 4.6.1 4.6.15g 4.6.15j 4.6.15n 4.6.1 |
|||
Group VIII (7 samples) 8/ Strain relief (static only) (UL 498) (3 samples) 9/ Strain relief (abrupt removal) (UL 498) (4 samples) 8/ Strain relief (abrupt removal) (UL 498) (4 samples) 9/ Visual and mechanical |
x x x |
x x x |
3.6.14o 3.6.14o 3.6.14o 3.1, 3.4, 3.5, 3.7 and 3.8 |
4.6.15o 4.6.15o 4.6.15o 4.6.1 |
1/ If single failure in group occurs, sample size of group to be doubled and entire test sequence repeated. Any additional failures disqualify product.
2/ Not required if UL or CSA listed.
3/ Isolated ground receptacle only.
4/ For 15 and 20 ampere, 125 volt duplex, nonlocking receptacles only.
5/ Only for receptacles with a U shape ground pin opening intended to receive 0.190 by 0.843-inch
(4.8260 by 21.412-mm) (max) plug grounding pin.
6/ Rubber and PVC materials only.
7/ Not required if UL or CSA listed as Hospital Grade.
8/ Angle plugs require an additional 8 samples.
9/ Not required for cable outlets if plugs in this group employ identical cord grip construction.